DNA Dot Blot Protocol
A. Buffers and Reagents
- 20X Saline Sodium Citrate (SSC) Buffer: 3.0 M NaCl, 0.3 M Sodium Citrate, pH to 7.0.
- 10X SSC Buffer: Dilute 20X SSC buffer 1:2.
- 2X DNA Denaturing Buffer: 200 mM NaOH, 20 mM EDTA.
- Nuclease-Free Water: (#12931)
- Blotting Membrane: This protocol has been optimized for positively charged nylon membranes.
- 96-Well Dot Blot Apparatus
- 10X Tris Buffered Saline with Tween® 20 (TBST): (#9997) To prepare 1 L 1X TBST: add 100 ml 10X TBST to 900 ml dH2 0, mix.
- Nonfat Dry Milk: (#9999)
- Blocking Buffer: 1x TBST with 5% w/v nonfat dry milk; for 150 ml, add 7.5 g nonfat dry milk to 150 ml 1X TBST and mix well.
- Bovine Serum Albumin (BSA): (#9998)
- Primary Antibody Dilution Buffer: 1X TBST with 5% BSA; for 20 ml, add 1.0 g BSA to 20 ml 1X TBST and mix well.
- Secondary Antibody Conjugated to HRP: anti-rabbit (#7074); anti-mouse (#7076).
- Detection Reagent LumiGLO® chemiluminescent reagent and peroxide (#7003) or SignalFire™ ECL Reagent (#6883)
B. Dot Blot
Note: This protocol is written for spotting fragmented, purified genomic DNA (titration of 1000 ng, 500 ng, 250 ng, 125 ng, 62.5 ng, 31.25 ng, and 15.625 ng) onto a positively charged nylon membrane using a 96-well dot blotting apparatus. Depending on the source and type of DNA, more or less DNA may be required for detection with the antibody.
Before Starting:
• Purify genomic DNA using a genomic DNA purification protocol or kit and sonicate
genomic DNA to generate fragments between 200 and 500 bp. DNA fragment size
can be analyzed by gel electrophoresis on a 1% agarose gel with a 100 bp DNA
marker.
• Cut a piece of nylon membrane to the size of the dot blot manifold.
• Wet nylon membrane with 10x SSC Buffer.
• Dry membrane by placing it in 96-well dot blot apparatus and applying vacuum.
-
Dilute fragmented genomic DNA to 100 ng/μl in 100 ul of nuclease-free water.
Then denature DNA by adding 100 μl of 2X DNA Denaturing Buffer and incubating at 95°C for 10 min. - Add 200 μl of 20X SSC buffer and immediately chill on ice for 5 min.
- Add 100 μl of nuclease-free water to bring DNA solution to a final volume of 500 μl with a DNA concentration of 20 ng/μl.
- Set up a series of six 2-fold dilutions by adding 250 μl of the DNA solution, starting with the DNA solution in Step 3, to 250 μl of nuclease-free water. This will generate seven DNA samples containing 250 μl DNA at concentrations of 20 ng/μl, 10 ng/μl, 5 ng/μl, 2.5 ng/μl, 1.25 ng/μl, 0.625 ng/μl, and 0.3125 ng/μl.
- Apply 50 μl of each of the seven dilution samples into separate wells of the 96-well dot blot apparatus, leaving the last well for nuclease-free water only. The amount of DNA added to each well should then be 1000 ng, 500 ng, 250 ng, 125 ng, 62.5 ng, 31.25 ng, 15.625 ng and 0 ng respectively. Apply gentle vacuum pressure to draw solution through the membrane. Nylon membrane should be mostly dry before step 6.
- Remove nylon membrane from the 96-well dot blot apparatus and wrap in plastic wrap.
- UV cross-link nylon membrane at 1200 J/m2.
C. Membrane Blocking and Antibody Incubation
- Incubate membrane in 25 ml of blocking buffer with gentle agitation for 1 hr at room temperature.
- Wash membrane three times for 5 min each with 15 ml of 1X TBST.
- Incubate membrane and primary antibody (at the appropriate dilution and diluent as recommended in the antibody product datasheet) in 10 ml primary antibody dilution buffer, with gentle agitation overnight at 4°C.
- Wash three times for 5 min each with 15 ml of 1X TBST.
- Incubate membrane with the species appropriate HRP-conjugated secondary antibody (#7074 Anti-rabbit IgG, HRP-linked Antibody or #7076 Anti-mouse IgG, HRP-linked Antibody) at 1:2000 in 10 ml of blocking buffer with gentle agitation for 1 hr at room temperature.
- Wash membrane three times for 5 min each with 15 ml of 1X TBST.
- Proceed with detection (Section D)
D. Detection of DNA
- Incubate membrane with 10 mL of LumiGLO® (0.5 ml 20x LumiGLO® #7003, 0.5 ml 20x Peroxide, and 9.0 ml purified water) or 10 ml SignalFire™ #6883 (5 ml Reagent A, 5 ml Reagent B) with gentle agitation for 1 min at room temperature.
- Drain membrane of excess developing solution (do not let dry), wrap in plastic wrap and expose to x-ray film. An initial 10 sec exposure should indicate the proper exposure time.
NOTE: Due to the kinetics of the detection reaction, signal is most intense immediately following incubation and declines over the following 2 hr
posted November 2015